The team conducted a detailed in-vivo study of Heron measurement repeatability to provide existing and prospective users with evidence about the instrument. Ten volunteers aged 19 to 40, all considered to have healthy corneal endothelium, were selected. Some had undergone LASIK or PRK, while others had no surgical history.
Three trained operators examined the volunteers’ 20 eyes, with ten examinations per eye. Across 200 measurements, mean endothelial density and standard deviation were calculated for each eye. The linked PDF contains the results table.
The report stated an overall standard deviation of 140 cells per square millimeter across all 200 measurements. It compared this with literature describing standard deviations up to 300 cells per square millimeter as acceptable. The team planned publication in a specialist journal after expanding the study population. These are historical study results and plans, rather than a new claim of clinical validation.